Intrinsically safe temperature probe with iTHERM QuickNeck for fast disassembly of the insert
Specs at a glance
Accuracy
Class AA acc. to IEC 60751 Class A acc. to IEC 60751 Class B acc. to IEC 60751 Class special or standard acc. to ASTM E230 Class 1 or 2 acc. to IEC 60584-2
Response time
fastest response time with thermowell t90 starting at below 10 s depending on configuration
Max. process pressure (static)
depending on the configuration up to 100 bar
Operating temperature range
PT100 TF iTHERM StrongSens: -50 °C ...500 °C (-58 °F ...932 °F) PT100 TF iTHERM QuickSens: -50 °C …200 °C (-58 °F …392 °F) PT100 WW: -200 °C ...600 °C (-328 °F ...1.112 °F) PT100 TF: -50 °C ...400 °C (-58 °F ...752 °F) Typ K: max. 1.100 °C (max. 2.012 °F) Typ J: max. 800 °C (max. 1.472 °F) Typ N: max. 1.100 °C (max. 2.012 °F)
iTHERM ModuLine TM131 is a modular metric RTD or TC thermometer with protection tube or to be used with onsite thermowell. It is suitable for every application, from basic functionality to high-end performance across all industries.
For universal use across all industries: From heavy industries (Power & Energy, Chemical, Oil & Gas) to utility applications.
International certification: e.g. explosion protection according to ATEX, IECEx, CSA C/US and NEPSI.
iTEMP temperature transmitter with all common communication protocols and optional Bluetooth® connectivity.
Benefits
Modular configuration for all applications: A wide selection of components and innovations, all combined under a single order code.
Increased efficiency: Fastest response times with QuickLink thermowell and optimal thermal coupling or iTHERM QuickSens sensor design.
Enhanced safety: Reliability in demanding applications with iTHERM StrongSens and highest vibration-resistance or Dual Seal second process barrier.
To meet the challenges of temperature monitoring in such critical processes, a thermowell design has been developed that responds to process changes up to five times faster than standard designs.
The Ziegler-Natta process uses hazardous chemicals like titanium tetrachloride. Precise temperature measurement is crucial to prevent runaway reactions and ensure safety and efficiency.
Torishima designs and builds large cooling water systems for customers in the Middle East, relying on Endress+Hauser instrumentation for the quality and performance of the products.
metric style modular temperature assembly universal range of application suitable for hazardous areas with neck or iTHERM QuickNeck incl. thermowell / protection tube (metal) or to be installed into thermowell can be used with iTHERM StrongSens, iTHERM QuickSens insert for fast response times
Thermowell / protection tube
welded protection tube or to be used with thermowell
Insert / probe
mineral insulated (MI), flexible
Outer diameter protection tube / Insert
Thermowell: 9x1,25 mm 11x2 mm 12x2,5 mm 14x2 mm 16x3,5 1/4" SCH80 1/2" SCH80 1/2" SCH40
straight reduced tapered optimized for quick response times
Surface roughness Ra
< 0.76 μm (0.03 μin)
Operating temperature range
PT100 TF iTHERM StrongSens: -50 °C ...500 °C (-58 °F ...932 °F) PT100 TF iTHERM QuickSens: -50 °C …200 °C (-58 °F …392 °F) PT100 WW: -200 °C ...600 °C (-328 °F ...1.112 °F) PT100 TF: -50 °C ...400 °C (-58 °F ...752 °F) Typ K: max. 1.100 °C (max. 2.012 °F) Typ J: max. 800 °C (max. 1.472 °F) Typ N: max. 1.100 °C (max. 2.012 °F)
Max. process pressure (static)
depending on the configuration up to 100 bar
Accuracy
Class AA acc. to IEC 60751 Class A acc. to IEC 60751 Class B acc. to IEC 60751 Class special or standard acc. to ASTM E230 Class 1 or 2 acc. to IEC 60584-2
Response time
fastest response time with thermowell t90 starting at below 10 s depending on configuration
Integration head transmitter
yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)
Ex - approvals
ATEX Ex ec, Ex tc ATEX IECEx Ex ta/tb, Ex ia, Ex db CSA C/US IS, NI, XP, DIP INMETRO Ex ia, Ex d, Ex tb NEPSI Ex ia, Ex d, Ex tD EAC Ex d, Ex ta/tb, Ex ia UKCA Ex ia, Ex nA, Ex tc KC Ex ia
Active barrier, 1 channel for 4 to 20 mA, HART® transparent with 24 to 230 V AC/DC as well as active/passive input and output, optionally with SIL and Ex
WirelessHART communication interface Transmission rate: Nominal 250 kBits/s Operating frequency: 2.4 GHz (ISM band) Transmission range under reference conditions: Outdoor 250 m, indoor 50 m
Active barrier, 1 channel for 4 to 20 mA, HART® transparent with 24 to 230 V AC/DC as well as active/passive input and output, optionally with SIL and Ex
WirelessHART communication interface Transmission rate: Nominal 250 kBits/s Operating frequency: 2.4 GHz (ISM band) Transmission range under reference conditions: Outdoor 250 m, indoor 50 m
Active barrier, 1 channel for 4 to 20 mA, HART® transparent with 24 to 230 V AC/DC as well as active/passive input and output, optionally with SIL and Ex